Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Experimental and Theoretical Characterization of an AC Electroosmotic Micromixer
Naoki SASAKITakehiko KITAMORIHaeng-Boo KIM
Author information
JOURNAL FREE ACCESS

2010 Volume 26 Issue 7 Pages 815-819

Details
Abstract

We have reported on a novel microfluidic mixer based on AC electroosmosis. To elucidate the mixer characteristics, we performed detailed measurements of mixing under various experimental conditions including applied voltage, frequency and solution viscosity. The results are discussed through comparison with results obtained from a theoretical model of AC electroosmosis. As predicted from the theoretical model, we found that a larger voltage (∼20 Vp-p) led to more rapid mixing, while the dependence of the mixing on frequency (1 − 5 kHz) was insignificant under the present experimental conditions. Furthermore, the dependence of the mixing on viscosity was successfully explained by the theoretical model, and the applicability of the mixer in viscous solution (2.83 mPa s) was confirmed experimentally. By using these results, it is possible to estimate the mixing performance under given conditions. These estimations can provide guidelines for using the mixer in microfluidic chemical analysis.

Content from these authors
© 2010 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback