Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Electrochemical Sensing of Acyclovir at a Gold Electrode Modified with 2-Mercaptobenzothiazole–[5,10,15,20-tetrakis-(3-methoxy-4-hydroxyphenyl)porphyrinato]copper(II)
Renjini JOSEPHK. Girish KUMAR
Author information
JOURNAL FREE ACCESS

2011 Volume 27 Issue 1 Pages 67-72

Details
Abstract
The electrooxidation of acyclovir (ACV) was studied using a gold electrode (GE) modified with a self-assembled monolayer of 2-mercaptobenzothiazol (MBZ) and [5,10,15,20-tetrakis(3-methoxy-4-hydroxyphenyl)porphyrinato]copper(II) (TMHPP Cu(II)) by square wave voltammetry (SWV). The self-assembled films are stable and showed blocking characteristics towards the faradaic processes such as gold surface oxidation and under potential deposition of copper. The optimized conditions obtained for the MBZ/TMHPP Cu(II)-modified GE were 0.1 M phosphate buffer solution (pH 7.0), square wave frequency of 15 Hz and square wave amplitude of 25 mV. Under these optimum conditions, the resultant peak current increases linearly with the concentration of ACV in the range of 1.0 × 10−3 to 1.0 × 10−8 M with a detection limit of 1.0 × 10−8 M. The MBZ/TMHPP Cu(II)-modified GE showed good stability and selectivity and it can be used to quantify ACV in pharmaceutical formulations and urine samples.
Content from these authors
© 2011 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top