Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Membrane Sampling with Microdialysis Coupled to HPLC/UV for On-line Simultaneous Determination of Melamine and Cyanuric Acid in Non-dairy Coffee Creamer
Yu-Ying CHAOYu-Tzu WEICheuch-Ting LEEHwang-Shang KOUYeou-Lih HUANG
Author information
JOURNAL FREE ACCESS

2011 Volume 27 Issue 10 Pages 1025

Details
Abstract
An on-line microdialysis/high-performance liquid chromatography method was developed for the simultaneous determination of melamine and cyanuric acid in non-dairy coffee creamer. To collect these analytes from aqueous samples, the microdialysis system featured a microdialysis probe incorporating a polyarylethersulfone membrane and employed 0.05 M HCl in 0.1% (v/v) MeOH as the perfusate, with optimal efficiency obtained at a flow rate of 1 μL min−1. The chromatographic conditions were optimized when using a reverse-phase phenyl column and a mobile phase of phosphate buffer solution in 10% (v/v) MeOH, buffered at pH 3.0. Good linearity relationship (r2 > 0.9987), intra- and inter-day precisions (RSDs < 6.6%), recoveries (96.9 – 105.0%), and limits of detection (melamine, 3 ppb; cyanuric acid, 150 ppb) were observed for the two analytes. This method has been successfully applied to simultaneous determination of melamine and cyanuric acid in commercial creamers with the recoveries in the range of 97.5 to 102.6%.
Content from these authors
© 2011 by The Japan Society for Analytical Chemistry
Previous article Next article