Mix-mode TiO-C18 Monolith Spin Column Extraction and GC-MS for the Simultaneous Assay of Organophosphorus Compounds and Glufosinate, and Glyphosate in Human Serum and Urine
Takeshi SAITOHiromichi AOKIAkira NAMERAHiroshi OIKAWAShota MIYAZAKIAkihiro NAKAMOTOSadaki INOKUCHI
Author information
JOURNAL FREE ACCESS

2011 Volume 27 Issue 10 Pages 999

Details
Abstract

A rapid, specific, and sensitive method for the simultaneous quantitation of organophosphates (fenitrothion (MEP), malathion, and phenthoate (PAP)), glufosinate (GLUF), and glyphosate (GLYP) in human serum and urine by gas chromatography–mass spectrometry (GC-MS) has been validated. All of the targeted compounds together with the internal standard were extracted from the serum and urine using a mix-mode TiO-C18 monolithic spin column. The recovery of organophosphates from serum and urine ranged from 12.7 to 49.5%. The recovery of GLUF and GLYP from serum and urine ranged from 1.9 to 7.9%. The intra- and inter-accuracy and precision (expressed as relative standard deviation, %RSD) were within 96.7 – 107.7% and 4.0 – 13.8%, respectively. The detection and quantitation limits for serum and urine were 0.1 and 0.1 μg/ml, respectively, for organophosphates, 0.1 and 0.5 μg/ml, respectively for GLUF and GLYP. The method had linear calibration curves ranging from 0.1 to 25.0 μg/ml for organophosphates and 0.5 – 100.0 μg/ml for GLUF, and GLYP. The validated method was successfully applied to a clinical GLYP poisoning case.