Counter Current “Emulsion Flow” Extractor for Continuous Liquid–Liquid Extraction from Suspended Solutions
Nobuyuki YANASEHirochika NAGANAWATetsushi NAGANOJunji NORO
Author information
JOURNAL FREE ACCESS

2011 Volume 27 Issue 3 Pages 325

Details
Abstract

A single current “emulsion flow” liquid–liquid extraction apparatus has a head with a number of holes from which micrometer-sized droplets of an aqueous phase spout into an organic phase to mix the two liquid phases. For practical use, however, a fatal problem can occur when particulate components in the aqueous phase plug the holes. In the present study, we have succeeded in solving the problem by applying a counter current-type emulsion flow extractor where micrometer-sized droplets of the organic phase are generated.

Content from these authors
© 2011 by The Japan Society for Analytical Chemistry