Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Evaluation of Non-aqueous Capillary Zone Electrophoresis for the Determination of Histamine H2 Receptor Antagonists in Pharmaceuticals
J. J. Berzas NEVADOG. Castañeda PEÑALVOR. M. Rodríguez DORADO
Author information
JOURNAL FREE ACCESS

2011 Volume 27 Issue 4 Pages 427

Details
Abstract

A new non-aqueous capillary electrophoresis method is proposed for the separation and simultaneous determination of cimetidine, ranitidine, roxatidine, nizatidine and famotidine by using a 30-cm long × 75 μm i.d. fused silica capillary and UV detection at 214 nm. Using a temperature of 25°C, an applied voltage of 15 kV, and a background electrolyte consisting of methanol containing 10 mM ammonium acetate and 0.2% acetic acid allowed the analytes to be separated in less than 4 min. The limits of detection obtained ranged from 7 and 17 μg L−1. The proposed method was successfully used to determine the analytes in pharmaceutical preparations and its results were checked against an HPLC method.

Content from these authors
© 2011 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback