Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Development of a N-Acetyl-β-D-glucosaminidase (NAG) Assay on a Centrifugal Lab-on-a-compact-disc (Lab-CD) Platform
Yoshihide TANAKASeira OKUDAAyumi SAWAIShigeo SUZUKI
Author information
JOURNAL FREE ACCESS

2012 Volume 28 Issue 1 Pages 33

Details
Abstract
A centrifugal microfluidic platform, which is also known as lab-on-a-compact-disc (Lab-CD), was developed for use as a urinary N-acetyl-β-D-glucosaminidase (NAG) activity assay. In this work, Lab-CD design, centrifugal operations and analytical procedures were established. Automated liquid handling on Lab-CD processes, such as fluid transport, sample metering, mixing, and fluorescence detection are accomplished using a portable Lab-CD system. The linearity of the NAG assay using 4-methylumbelliferyl-N-acetyl-β-D-glucosaminide (4-MU-GlcNAc) was found to be acceptable in the range of 2.5 to 20 U L−1; the relative standard deviations for the fluorescence intensity of eight samples (7.5 U L−1) was 6.4%. Clinical diagnostics is one of the most promising applications for Lab-CD technologies. All the benefits of miniaturization, such as reduced sample requirement, reduced reagent consumption and automation, are realized in this investigation.
Content from these authors
© 2012 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top