Simultaneous Speciation of Iron(II) and Iron(III) by Ion Chromatography with Chemiluminescence Detection
Yun-Chieh CHENYu-Ling JIANKong-Hwa CHIUHwa-Kwang YAK
Author information
JOURNAL FREE ACCESS

2012 Volume 28 Issue 8 Pages 795-799

Details
Abstract
This study reports on a method for the speciation of iron in aqueous samples by the simultaneous analysis of divalent and trivalent iron ions with ion chromatography equipped with chemiluminescence detection (IC-CLD). Ferrous and ferric ions are first chelated by pyridine-2,6-dicarboxylic acid (PDCA) to form complexed anions, and separated by a mixed-bed ion-exchange column. The separated complexed ions are then detected with a CLD system containing luminol and hydrogen peroxide in a basic solution. This luminescence system has a linear dynamic range of ca. 3 orders of magnitude, with method detection limits as low as 7 μg L−1 for Fe(II) and 3 μg L−1 for Fe(III), measured in the simultaneous detection mode. This system resists interferences from common cations such as Cd, Ca, Cr, Cu, Mg, Ni, Pb, and Zn. Evaluation by analyzing real samples shows that this method is rapid, accurate, sensitive, and selective.