Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Determination of Glycoproteins by a Self-Assembled 4-Mercaptophenylboronic Acid Film on a Quartz Crystal Microbalance
Xiao-Tong MAXi-Wen HEWen-You LIYu-Kui ZHANG
Author information
JOURNAL FREE ACCESS

2016 Volume 32 Issue 12 Pages 1277-1282

Details
Abstract

Glycosylation plays an important part in many biological processes. However, many glycoproteins are either of low abundance or covered by other components in biological samples. Hence, developing new methods to measure the glycoproteins with both high efficiency and low detection limit is important. In this work, a self-assembled 4-mercaptophenylboronic acid film was coated on a quartz crystal microbalance chip. By optimizing the reaction time and the concentration of 4-mercaptophenylboronic acid, a sensor that specifically responded to glycoproteins was created. Then, several parameters for the prepared sensor were investigated and the working curve for representative glycoprotein–transferrin was established. The linearity range was from 50 to 400 ng/mL and the detection limit was 21.0 ng/mL. The sensor was used to detect transferrin in artificial urine samples. This sensor has a low detection limit of glycoproteins requiring only a small amount of samples, and thus has potential applications in both pharmaceutical and medical areas.

  Fullsize Image
Content from these authors
© 2016 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top