Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Microfluidic Paper-based Analytical Device for the Determination of Hexavalent Chromium by Photolithographic Fabrication Using a Photomask Printed with 3D Printer
Hitoshi ASANOYukihide SHIRAISHI
Author information
JOURNAL FREE ACCESS

2018 Volume 34 Issue 1 Pages 71-74

Details
Abstract
This article describes a simple and inexpensive microfluidic paper-based analytical device (μPAD) for the determination of hexavalent chromium (CrVI) in water samples. The μPADs were fabricated on paper by photolithography using a photomask printed with a 3D printer and functionalized with reagents for a colorimetric assay. In the μPAD, CrVI reacts with 1,5-diphenylcarbazide to form a violet-colored complex. Images of μPADs were captured with a digital camera; then the red, green, and blue color intensity of each detection zone were measured using images processing software. The green intensity analysis was the best sensitive among the RGB color. A linear working range (40 – 400 ppm; R2 = 0.981) between the CrVI and green intensity was obtained with a detection limit of 30 ppm. All of the recoveries were between 94 and 109% in recovery studies on water samples, and good results were obtained.
Fullsize Image
Content from these authors
© 2018 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback