Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Amperometric Determination of Hydrogen Peroxide in Whitening Gels Using Boron-doped Diamond Electrode
Gustavo C. AZEVEDORoberto H. S. CASTROMaria Auxiliadora C. MATOSRenato Camargo MATOS
Author information
JOURNAL FREE ACCESS

2018 Volume 34 Issue 8 Pages 913-917

Details
Abstract

The aim of this work was to develop an electrochemical cell and a methodology based on an amperometric determination of hydrogen peroxide in whitening gel samples under a boron-doped diamond electrode using flow injection analysis. Different parameters were evaluated to obtain the best conditions of analysis: among them, the flow of electrolyte at 2.8 mL min−1, the loop sampling 175 μL (28.5 cm), an analytical length of 159 μL (25 cm) and an applied potential of +0.60 V vs. Ag/AgCl(sat). The proposed method was suitable in terms of precision of results (RSD <10%); the accuracy was confirmed in the analysis of the gels through addition and recovery studies with results between 74 and 107%. The method was then applied to the analysis of tooth-whitening gel samples, acquired in different cities of the region. Regarding the results, a medium concentration value of 2.39% (w/w) was observed.

Fullsize Image
Content from these authors
© 2018 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top