Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Determination of Ultra-trace Tellurium in Steel by ID-ICP-MS/MS with Liquid–Liquid Extraction
Junichi HIRATADaisuke ITABASHIMichihiro AIMOTO
Author information
JOURNAL FREE ACCESS

2021 Volume 37 Issue 12 Pages 1757-1763

Details
Abstract

Inductively coupled plasma mass spectrometry using isotope dilution (ID-ICP-MS) with liquid–liquid extraction was used for determining ultra-trace tellurium (Te) in steels to improve the sensitivity, accuracy, and precision of the analysis. Single quadrupole-type ICP-MS cannot be used to determine the trace amount of Te because of the contribution of mass spectrometric interference by Xe. To overcome this, tandem mass-filter-type ICP-MS (ICP-MS/MS) was used to determine Te. A pretreatment by liquid–liquid extraction was also employed to compensate for the decrease in the signal intensity due to the use of ICP-MS/MS. The relative standard deviations of the resulting abundances, defined by repeated analysis from separated three steel samples, were <1.3%. Furthermore, the determined values of Te in standard reference materials were close to the reference values. Thus, the developed determination method is useful for the analysis of ultra-trace Te in steels.

Fullsize Image
Content from these authors
© 2021 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top