Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Flow Injection Analysis for Trace Aluminum with On-Line Preconcentration and Spectrophotometric Detection
Jun KOBAYASHIMiyako BABAMotoichi MIYAZAKI
Author information
JOURNAL FREE ACCESS

1994 Volume 10 Issue 2 Pages 287-291

Details
Abstract

We developed a flow injection analysis method for the determination of trace aluminum in water coupled with on-line preconcentration on salicylideneamino-2-thiophenol immobilized glass beads (SAB) and spectrophotometric detection using Pyrocatechol Violet (PV). The FIA system consisted of two pumps, a 6-port switching valve, a concentrator column, a UV/VIS spectrophotometric detector and a recorder. A sample solution was combined with a succinic acid/ NaOH buffer (pH 6.0) and hydroxylammonium chloride and this mixture was loaded on the concentrator column. Aluminum retained on the concentrator column was eluted with 0.1mol dm-3 nitric acid and mixed with PV. A colored PV-aluminum complex was spectrophotometrically monitored at 576nm. A linear relationship was observed between the peak height and aluminum concentration at a constant sample volume. A correlation coefficient of 0.9997 and determination limit of 3.7nmol dm-3 were obtained for a sample volume of 10ml. Relative standard deviations were below 2% at nmol levels. Aluminum traces in some Kanazawa City tap water samples were determined by this method.

Content from these authors
© The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top