Optimization of Operating Conditions in Individual Airborne Particle Analysis by Inductively Coupled Plasma Mass Spectrometry
Satoshi KANECOTsutomu NOMIZUTomokazu TANAKANaoto MIZUTANIHiroshi KAWAGUCHI
Author information
JOURNAL FREE ACCESS

1995 Volume 11 Issue 5 Pages 835-840

Details
Abstract

Operating conditions of a particle analyzer using inductively coupled plasma mass spectrometry were investigated to optimize the analysis of individual aerosol particles for zinc in the picogram to femtogram range. The signal intensities and their fluctuations from individual particles were strongly affected by the carrier gas flow rate, radio frequency power and sampling depth. The optimum operating conditions were determined and are discussed based on the vaporization and ionization processes of particles and the ion sampling process. Under the optimum operating conditions, fluctuations of the signals from monodisperse aerosols were improved to ca. 5% of relative standard deviation under favorable conditions, and the detection limit of zinc in a particle could be lowered to ca. 3fg.