Flow Injection Titration of Chloride in Electroplating Baths With a Silver Tubular Electrode as a Detector
Begoña ETXEBARRIAJos L. F. C. LIMAConceição B. S. M. MONTENEGRORicardo REZ-OLMOS
Author information
JOURNAL FREE ACCESS

1997 Volume 13 Issue 1 Pages 89-92

Details
Abstract
A flow injection system has been constructed for the "pseudo-titration" of chloride in electroplating baths with potentiometric detection. A silver ion-selective tubular electrode based on a homogeneous crystalline membrane was used. The FIA manifold has been developed and optimized in order to analyze samples with a chloride concentration range of 6.O×10-2M to 2.5M. The proposed method is precise (relative standard deviation about 5%) and the results obtained by a simultaneous application of the FIA method and an automatic precipitation titration method, adopted as a reference technique, are in good agreement (r=0.996). The sampling rate for the FIA manifold is 78 samples per hour.