Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Determination of Pheophorbide a and Methyipheophorbide a by Fluorescence High Performance Liquid Chromatography: Availability of Narrow Baseline Method
Yoshiaki YOSHITAKEKikuko KAKIUCHIKiyotoshi MORISHIGEShiro GOHDAYasuharu NISHIKAWATsunenobu SHIGEMATSU
Author information
JOURNAL FREE ACCESS

1986 Volume 2 Issue 1 Pages 67-69

Details
Abstract

A narrow baseline method for the simultaneous determination of pheophorbide a and methylpheophorbide a in core sediment samples has been developed by high performance liquid chromatography. A linear relationship between the concentration of methylpheophorbide a and the peak height on the narrow baseline was obtained. The relative standard deviation of measurement was found to be 2.4% (n=6) for samples containing 8.1×10-8mol dm-3 ofmethylpheophorbide a.