Effects of Ion Optics on the Sensitivity Factors in Secondary Neutral Mass Spectrometry Measurements
Yoshikazu KIKUTAHolger JENETT
Author information
JOURNAL FREE ACCESS

1991 Volume 7 Issue 5 Pages 757-761

Details
Abstract
The effects of the ion optics of the SNMS (secondary neutral mass spectrometry) INA-3 system on the intensity ratios, from which the relative sensitivity factors can be calculated, have been investigated for several III-V semiconductors: GaAs, GaP, and InP. It was found that the intensity ratios change significantly, depending on the electrode potentials of the ion optics and on the mass difference (M) between the major constituent elements: within 90% regions of optimum electrode voltages and between M=6 and 84amu, the maximum total relative deviation of relative intensities increases from about 10% to 250%.