2024 Volume 10 Issue 1 Pages 59-63
Abstract
Despite recent developments in positron emission tomography technology, cardiac single photon emission computed tomography (SPECT) imaging continues to be the main stream of nuclear cardiology because of its high accessibility and wider clinical use. For SPECT imaging, cadmium-zinc telluride (CZT) based detectors have advantages over conventional Anger type detectors with more flexible camera design thanks to the small sized CZT detectors and high contrast imaging because of its high energy resolution. Approximately 15 years ago, CZT-based cardiac SPECT cameras became commercially available, resulting in great success in clinical practice. However, this was just the beginning of the CZT era. To date, CZT-based general purpose whole-body camera and 3 dimensional full-ring SPECT systems have been developed and become commercially available with promising initial results. Although there still are areas to be addressed before wider clinical use, the CZT-based technology may provide significant progresses in nuclear cardiology as new generation SPECT systems. Finally, newer materials for the semiconductor detectors are under investigation or development, suggesting that there will be more developments in cardiac SPECT technology.
Despite recent developments in positron emission tomography (PET) technology, cardiac single photon emission computed tomography (SPECT) imaging continues to be the main stream of nuclear cardiology because of its high accessibility and wider clinical use. For SPECT imaging, cadmium-zinc telluride (CZT) based detectors have advantages over conventional Anger type detectors. These include that 1) the small sized CZT detector enables flexible camera design such as a ring-configured SPECT system resulting in high sensitivity, which is achieved by direct conversion of gamma ray to electric signal, and high spatial resolution, which is achieved by small detector pixel size (2.46 mm), and that 2) the high energy resolution enables high contrast imaging. Therefore, CZT detectors have been applied to many SPECT imaging devices such as pre-clinical SPECT system, SPECT camera dedicated for cardiac imaging, whole-body SPECT system, and 3 dimensional (3D) full-ring SPECT system. The purpose of this review article was to review and discuss about various CZT-based SPECT systems in view of clinical utility and future perspectives.
Pre-clinical SPECT
It is conceivable that CZT detectors are especially suitable for pre-clinical imaging because of its small detector size. In fact, a couple of vendors developed such CZT-based SPECT systems. Figure 1 depicts a CZT-based SPECT/CT system (eXplore speCZT,GE Healthcare, Milwaukee, WI) designed for small animal imaging such as rats and mice (1). Using this system, electrocardiogram (ECG)-gated myocardial β-methyl iodophenyl-pentadecanoic acid (BMIPP) imaging in the rat was feasible as shown in Figure 2.
Cardiac SPECT
For clinical use, CZT technology was applied to SPECT systems dedicated for cardiac imaging. Currently, two SPECT systems are commercially available. One is MyoSPECT (ex. Discovery MM 530c (2), GE Healthcare, Milwaukee, WI) and the other is D-SPECT (3) (Spectrum Dynamics, Israel). Figure 3 depicts a CZT-based cardiac SPECT with stationary detectors with a pin-hole collimator (Discovery NM 530c). Both SPECT systems have advantages over conventional SPECT in two ways. Firstly, fast imaging with short acquisition time or low dose imaging for low radiation exposure is feasible without loss in diagnostic performance as demonstrated in many studies (4–6). This was true when compared to diagnostic performance using conventional Anger SPECT (7). Secondly, myocardial flow reserve can be estimated by dynamic SPECT imaging capability (8, 9). A limitation of flow reserve measurement using SPECT is, however, the lack of ideal Tc-99m labeled flow tracers particularly for high flow rates. Because of this, accurate measurements of hyperemic flow is limited as compared to PET. Nevertheless, clinical significance of CZT-based cardiac SPECT is now well acknowledged and accepted in clinical practice.
Whole-body SPECT
With the success of CZT-based cardiac SPECT systems, a CZT-based camera for general purpose whole-body imaging has been developed. Discovery NM/CT 670 CZT (currently, NM/CT 870 CZT, GE Healthcare, Milwaukee, WI) is the only commercially available CZT-based camera for whole-body imaging. As illustrated in Figure 4, this system has 2 large field of view CZT-based gamma cameras equipped with 16-raw CT scanner.
Using this system, we tested whether fast planar imaging is feasible without loss in diagnostic performance, because this is the only CZT-based camera with true planar imaging capability. As expected, fast 123I-MIBG planar imaging of as fast as 50 second acquisition was feasible without significant errors in measured heart-to-mediastinum uptake ratios (10). Similarly, fast whole-body bone imaging was also feasible (11). For cardiac tomographic imaging, fast ECG-gated myocardial perfusion SPECT as fast as 2 minute acquisition was feasible without significant loss in diagnostic performance, as shown in Figure 5 (12).
Another advantage of CZT detectors is its high energy resolution (5–6 % full-width at half maximum [FWHM]) as compared with conventional Anger detectors (8–10 % FWHM). On the basis of its high energy resolution, we tested the feasibility of simultaneous 99mTc-tetrofosmin (peak energy of 140 KeV) and 123I-BMIPP (peak energy of 159 KeV) dual-isotope imaging in patients with acute myocardial infarction (13). A representative case example of 99mTc-tetrofosmin/123I-BMIPP dual-isotope imaging is shown in Figure 6. As expected, a clear mismatch of myocardial distribution between the two tracers was observed, suggesting the feasibility of simultaneous dual-isotope imaging even when the energy peaks of two tracers are close to each other.
3D full-ring SPECT
Recent advances in technology enabled development of CZT-based 3D full-ring SPECT systems, which was achieved because of small sized CZT detectors. Two such systems (StarGuide, GE Healthcare, Milwaukee, WI and Veriton, Spectrum Dynamics, Israel) are currently commercially available. Both systems have 12 movable small detectors surrounding FOV. This configuration has three advantages over conventional two detector SPECT systems (Figure 7). Firstly, higher sensitivity can be achieved thanks to more detectors than conventional systems (12 versus 2). Secondly, higher spatial resolution can be obtained because detectors can be more close to the target body. Finally, dynamic SPECT imaging is feasible to obtain, for example, myocardial flow reserve.
Figure 8 depicts one of such 3D full-ring SPECT systems (StarGuide, GE Healthcare). As aforementioned, this system consists of 12 movable small detectors for SPECT imaging together with a 16-raw CT. The detailed system performance has been reported elsewhere (14). According to that report, the system has the energy resolution of 5.3% FWHM for 99mTc, the volume sensitivity of 520 cps/MBq, and SPECT spatial resolution of 4 mm FWHM in all directions. Additionally, unlike SPECT cameras dedicated for cardiac imaging, attenuation correction is ready to perform owing the combined CT system. An example of attenuation corrected myocardial SPECT images using this system are shown in Figure 9.
A disadvantage of 3D full-ring SPECT is that planar imaging is not possible because its configuration is dedicated for SPECT imaging. This is not negligible because heart-to-mediastinum uptake ratios on 123I-MIBG images, for example, is an important prognostic biomarker in heart failure patients (15). However, planar-like images (i.e., SPECT derived planar images) can be generated from projection data. An example of the SPECT derived planar image of 123I-MIBG phantom (16) is shown in Figure 10.
Thus, CZT-based 3D full-ring SPECT is promising with high system performance and may provide significant progresses in nuclear cardiology as a totally new generation SPECT system. However, there are areas to be addressed before its wide clinical acceptance. Firstly, available clinical data are still limited because this is a rather new system. Secondly, softwares for data acquisition and processing such as those for dynamic imaging should be refined.
Beyond CZT-based detectors
With the great success of CZT-based SPECT cameras, newer materials for the semiconductor solid-state detector are under investigation or development. Of these, thallium bromide seems to be a promising candidate with high sensitivity and high energy resolution (17). Thus, we hope that such newer materials for the detectors will become clinically available in near future.
Conclusion
CZT technology is one of the most exciting developments in nuclear cardiology field. Approximately 15 years ago, CZT-based cardiac SPECT cameras became commercially available, resulting in great success in clinical practice. However, this was just the beginning of the CZT era. To date, CZT-based general purpose whole-body camera and 3D full-ring SPECT systems have been developed and become commercially available with promising initial results. Although there still are areas to be addressed before wider clinical use, the CZT-based technology may provide significant progresses in nuclear cardiology as new generation SPECT systems. Finally, newer materials for the semiconductor detectors are under investigation or development, suggesting that there will be more developments in cardiac SPECT technology.
Acknowledgments
None.
Sources of funding
None.
Conflicts of Interest
The author has no conflict of interest.
References
1. Matsunari I, Miyazaki Y, Kobayashi M, et al. Performance evaluation of the eXplore speCZT preclinical imaging system. Ann Nucl Med 2014; 28: 484–97.
2. Esteves FP, Raggi P, Folks RD, et al. Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: Multicenter comparison with standard dual detector cameras. J Nucl Cardiol 2009; 16: 927–34.
3. Erlandsson K, Kacperski K, van Gramberg D, Hutton BF. Performance evaluation of D-SPECT: A novel SPECT system for nuclear cardiology. Phys Med Biol 2009; 54: 2635–49.
4. Gimelli A, Bottai M, Giorgetti A, Genovesi D, Filidei E, Marzullo P. Evaluation of ischaemia in obese patients: Feasibility and accuracy of a low-dose protocol with a cadmium-zinc telluride camera. Eur J Nucl Med Mol Imaging 2012; 39: 1254–61.
5. Gimelli A, Liga R, Bottai M, et al. Diastolic dysfunction assessed by ultra-fast cadmium-zinc-telluride cardiac imaging: Impact on the evaluation of ischaemia. Eur Heart J Cardiovasc Imaging 2015; 16: 68–73.
6. Herzog BA, Buechel RR, Husmann L, et al. Validation of CT attenuation correction for high-speed myocardial perfusion imaging using a novel cadmium-zinc-telluride detector technique. J Nucl Med 2010; 51: 1539–44.
7. Buechel RR, Herzog BA, Husmann L, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur J Nucl Med Mol Imaging 2010; 37: 773–8.
8. Agostini D, Roule V, Nganoa C, et al. First validation of myocardial flow reserve assessed by dynamic 99mTc-sestamibi CZT-SPECT camera: Head to head comparison with 15O-water PET and fractional flow reserve in patients with suspected coronary artery disease. The WATERDAY study. Eur J Nucl Med Mol Imaging 2018; 45: 1079–90.
9. Otaki Y, Manabe O, Miller RJH, et al. Quantification of myocardial blood flow by CZT-SPECT with motion correction and comparison with 15O-water PET. J Nucl Cardiol 2021; 28: 1477–86.
10. Okano N, Osawa I, Tsuchihashi S, Takahashi M, Niitsu M, Matsunari I. High-speed scanning of planar images showing 123I-MIBG uptake using a whole-body CZT camera: A phantom and clinical study. EJNMMI Res 2019; 9: 22.
11. Yamane T, Kondo A, Takahashi M, et al. Ultrafast bone scintigraphy scan for detecting bone metastasis using a CZT whole-body gamma camera. Eur J Nucl Med Mol Imaging 2019; 46: 1672–7.
12. Yamada Y, Nakano S, Gatate Y, et al. Shortened acquisition time in simultaneous 99mTc-tetrofosmin and 123I-β-methyl-p-iodophenyl pentadecanoic acid dual-tracer imaging with cadmium-zinc-telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction. Nucl Med Commun 2019; 40: 1130–7.
13. Yamada Y, Nakano S, Gatate Y, et al. Feasibility of simultaneous 99mTc-tetrofosmin and 123I-BMIPP dual-tracer imaging with cadmium-zinc-telluride detectors in patients undergoing primary coronary intervention for acute myocardial infarction. J Nucl Cardiol 2021; 28: 187–95.
14. Zorz A, Rossato MA, Turco P, et al. Performance evaluation of the 3D-ring cadmium-zinc-telluride (CZT) StarGuide system according to the NEMA NU 1-2018 standard. EJNMMI Phys 2024; 11: 69.
15. Nakata T, Nakajima K, Yamashina S, et al. A pooled analysis of multicenter cohort studies of 123I-mIBG imaging of sympathetic innervation for assessment of long-term prognosis in heart failure. JACC Cardiovasc Imaging 2013; 6: 772–84.
16. Nakajima K, Okuda K, Matsuo S, et al. Standardization of metaiodobenzylguanidine heart to mediastinum ratio using a calibration phantom: Effects of correction on normal databases and a multicentre study. Eur J Nucl Med Mol Imaging 2012; 39: 113–9.
17. Ariño-Estrada G, Du J, Kim H, et al. Development of TlBr detectors for PET imaging. Phys Med Biol 2018; 63: 13NT04.