Annals of Nuclear Cardiology
Online ISSN : 2424-1741
Print ISSN : 2189-3926
ISSN-L : 2189-3926
Review Articles
Optimal Protocol of Myocardial Perfusion Imaging for Reduction of Radiation Exposure
Yasuyuki SuzukiNaoya MatsumotoShunichi YodaYasuo AmanoYasuo Okumura
Author information
JOURNAL FREE ACCESS FULL-TEXT HTML

2022 Volume 8 Issue 1 Pages 109-112

Details

Abstract

The Japanese Circulation Society (JCS) 2021 guideline on radiation safety in cardiology was updated based on the eight best practices for myocardial perfusion single-photon emission computed tomography myocardial perfusion imaging (SPECT-MPI), which was determined by panels of international atomic energy agency (IAEA) nuclear cardiology protocols cross-sectional study (INCAPS). Although the guideline recommends to avoid too much 201Tl or 201Tl/99mTc dual-isotope protocol, the utilization rate of 201Tl for SPECT-MPI in Japan is considerably high compared to other countries. In Japan, protocols using 99mTc should be appropriately promoted based on the guideline to reduce radiation exposure in SPECT-MPI. Among the eight practices, stress-only imaging is considered to be a useful strategy that contributes to a significant reduction in radiation exposure, and hybrid assessment in combination with rest computed tomography (CT)-MPI is one of the practical application methods.

Single-photon emission computed tomography myocardial perfusion imaging (SPECT-MPI) requires 201Tl or 99mTc labeled radiopharmaceuticals. Otsuka et al. reported that the utilization rate of 201Tl and 99mTc for SPECT-MPI in Japan in 2016 were almost the same, about 50% (1). Previous study reported that the rates of 201Tl protocol for SPECT-MPI in Germany and the USA were less than 10% (2–4). The rates of 201Tl for SPECT-MPI in Japan is considerably higher than that in other countries, and it seems to affect the increase of radiation exposure. International atomic energy agency (IAEA) nuclear cardiology protocols cross-sectional study (INCAPS) described the definition of eight best practices for reduction of the patients’ radiation exposure (5). In the Japanese Circulation Society (JCS) 2021 guideline on Radiation safety in Cardiology, optimization methods for SPECT-MPI were proposed based on the eight best practices. In this paper, the eight best practices are explained, and additional explanations are given on the stress only protocol as part of hybrid assessment in clinical scenario of sequential assessment (6, 7), which can significantly reduce radiation exposure even with conventional SPECT scanner, and camera-based dose-reduction strategies using semiconductor scanner.

The eight best practices

The following criteria were determined by a committee of expert physicians and medical physicists convened by IAEA. The core elements identified by the panel based on previous guidelines (8, 9) included practices such as avoiding administering too much isotope, avoiding higher dose isotopes, and using newer technologies that can lower radiation doses.

1. Avoid 201Tl

The JCS 2021 guideline on radiation safety in cardiology recommended to avoid 201Tl stress tests in patients ≤ 70 years old. The INCAPS reported that 92% of participating institutions had met this standard (5). Previous study from Einstein et al. reported that SPECT-MPI performed with 201Tl is associated with a considerably higher radiation dose to patients than when it is performed with 99mTc (10). Note that this excludes 201Tl rest-redistribution viability studies and stress-redistribution-reinjection stress-and-viability studies.

2. Avoid dual-isotope

Dual-isotope (rest 201Tl and stress 99mTc) stress tests recommended to be avoided in patients ≤70 years old. Dual-isotope SPECT-MPI is associated with the highest radiation dose of any protocols (10). The INCAPS reported that 97% of participating institute had met this standard (5).

3. Avoid too much 99mTc

It is recommended to avoid 99mTc injection with administered activity >1,332 MBq (36 mCi), and mean total effective dose ≥ 15 mSv for all studies using 99mTc injections. 1,332 MBq is the highest recommended activity in guidelines (9), and 15 mSv is a high radiation dose for a study using 99mTc. This standard is achieved in Japan by complying with the diagnostic reference level (DRL; 1,220 MBq) in Japan.

4. Avoid too much 201Tl

For each nuclear stress test involving 201Tl, no more than 129.5 MBq (3.5 mCi) should be administered at stress. The expert committee maintained that no more than this activity is needed for patients who are good candidates to receive 201Tl SPECT-MPI. This standard was achieved in 99% of INCAPS participating institutions (5).

5. Perform stress-only imaging

If stress images are completely normal, subsequent rest imaging can be avoided in order to reduce radiation dose up to 80% (5). In previous reports on stress-only imaging, rest testing was omitted in 2/3 of all examinations, which is considered to greatly contribute to the reduction of radiation exposure (11). The global average rate of stress-only imaging was 30%, indicating a particularly low trend (5). In stress-only imaging, the need for additional resting imaging must be determined reliably. In the clinical scenario of sequential imaging in which coronary computed tomography (CT) is performed prior to SPECT-MPI, stress-only imaging should be considered more actively to enable hybrid assessment, which will be discussed later.

6. Use camera-based dose reduction strategies

Following dose reduction strategies are recommended: (a) attenuation correction (CT or line source), (b) imaging patients in multiple positions (e.g. both supine and prone), (c) high-technology software (e.g. incorporating iterative reconstruction, resolution recovery, and noise reduction), and (d) high-technology hardware (e.g. a high-efficiency solid state SPECT camera or cardiac-focused collimator). The JCS 2021 guideline recommended longer scan time to reduce radiation dose instead of CT attenuation collection. Each of the approaches reduces the radiation dose and image noise, thereby facilitating stress-only imaging.

7. Weight-based dosing for 99mTc

In Japan, most of the study site use syringe kit as a type of radiopharmaceuticals which is difficult to adjust weight-based dosing. The JCS guideline recommended to reduce the applied dose in patients < 50 kg.

8. Avoid inappropriate dosing that can lead to ‘shine through’ artifact

Shine through occurs in 99mTc one-day protocol. To avoid shine through, it is recommended that the radioactivity imaged for the second injection be at least three to four times that of the first injection. Relatively short half-life 99mTc agent contributes to the reduction of radiation exposure in the second injection of 99mTc agent. It should be noted that INCAPS described that avoiding ‘shine through’ does not directly resulted in radiation dose reduction. According to the hierarchical regression model, avoiding ‘shine-through’ resulted in 1.03 mSv “increase” of effective dose (5).

Hybrid assessment using stress-only imaging

Stress-only imaging requires reliable assessment of the need for additional resting images for all stress images, which may be a barrier to implementation. If stress-only imaging was normal, no more resting images are required. Conversely, stress-only imaging with perfusion defect needs to have resting images for eliminating myocardial infarction or artifacts. Sequential assessment with SPECT-MPI after anatomical assessment of the coronary arteries using coronary CT with judgement of intermediate coronary stenosis or insufficient image quality is considered an appropriate clinical choice (7). The image dataset from coronary CT can be used not only for anatomical assessment of the coronary arteries, but also for resting myocardial perfusion, including myocardial scar or infarction detection. Gerber et al. reported that CT perfusion imaging (CT-MPI) showed good enough correlation with delayed enhancement on magnetic resonance imaging for assessment of the size of the myocardial infarction (12). Therefore, in the clinical scenario of sequential assessment, image dataset of CT-MPI can be substituted for rest SPECT-MPI in addition to stress-only imaging. Suzuki et al. proposed hybrid assessment, which combines stress SPECT-MPI and rest CT-MPI, in sequential assessment as a practical application of stress-only imaging (6). Graphical explanation of hybrid assessment using stress-only SPECT-MPI and rest CT-MPI is shown in Figure 1. Suzuki et al. reported that total effective radiation dose of hybrid assessment with coronary CT was 4.6 mSv when administering 296 MBq of 99mTc-tetrofosmin, which was less than half of that using standard stress/rest SPECT-MPI with coronary CT (9.7 mSv), whereas diagnostic performance of hybrid assessment was not significantly different (conversion factor for 99mTc-tetrofosmin at stress (0.0069 mSv/MBq) and conversion factor of dose length product for the calculation of radiation dose of coronary CT(0.014 mSv/[mGy/cm]) were applied based on the previous recommendation) (6, 13). It should be noted that conversion factor for 99mTc-sestamibi at stress is slightly higher (0.0079 mSv/MBq) than that of 99mTc-tetrofosmin.

Hybrid assessment using stress-only imaging can be applied combination of other dose reduction strategies including weight-based and camera-based methods. Even when using a conventional gamma camera, stress-only imaging with a smaller amount of 99mTc can be achieved under some conditions. Stress-only imaging with 296 MBq of 99mTc can be performed with a conventional gamma camera in a patient with body mass index ≤ 25 (kg/m2) (14). Previous study from Zafrir et al. reported half-dose (185 MBq) 99mTc protocol for stress SPECT-MPI using a conventional gamma camera and dedicated image processing (15). Hybrid assessment with camera-based dose reduction strategies may lead lower level of radiation exposure than previously reported. The imaging guideline of the American Society of Nuclear Cardiology in 2016 recommended 148–222 MBq of 99mTc tracer for stress-only imaging if newer technology, including semiconductor scanners, is available (14). In addition, previous studies reported the potential of semiconductor scanners to achieve an extremely lower dose protocol using <148 MBq of 99mTc without sacrificing image quality or diagnostic performance (16, 17).

Conclusions

New guideline on radiation safety in cardiology in Japan was updated based on the eight best practices, which is important elements for reduction of patients’ radiation exposure. In Japan, protocols using 99mTc should be appropriately promoted based on the guideline to reduce radiation exposure in SPECT- MPI. In addition, among the eight best practices, stress-only imaging is considered to be a useful strategy that contributes to a significant reduction in radiation exposure, and hybrid assessment in combination with rest CT-MPI is one of the practical application methods.

Acknowledgments

The authors would like to thank radiology technicians Yoshitaka Hori and Hiroyuki Matusno for useful discussion and suggestion on the topic treated in this paper.

Sources of funding

None.

Conflicts of interest

None.

References

1. Otsuka R, Kubo N, Miyazaki Y, Kawahara M, Takaesu J, Fukuchi K. Current status of stress myocardial perfusion imaging pharmaceuticals and radiation exposure in Japan: Results from a nationwide survey. J Nucl Cardiol 2017; 24: 1850–5.

2. Becker MD, Butler PF, Bhargavan-Chatfield M, Harkness BA, Metter D, MacFarlane CR, et al. Adult gamma camera myocardial perfusion imaging: diagnostic reference levels and achievable administered activities derived from ACR accreditation data. J Am Coll Radiol 2016; 13: 688–95.

3. Songy B, Guernou M, Lussato D, Queneau M, Geronazzo R. Low-dose thallium-201 protocol with a cadmium-zinc-telluride cardiac camera. Nucl Med Commun 2012; 33: 464–9.

4. Lindner O, Burchert W, Schäfer W, Hacker M. Myocardial perfusion SPECT 2015 in Germany. Results of the 7th survey. Nuklearmedizin 2017; 56: 31–8.

5. Einstein AJ, Pascual TNB, Mercuri M, Karthikeyan G, Vitola JV, Mahmarian JJ, et al. Current worldwide nuclear cardiology practices and radiation exposure: results from the 65 country IAEA Nuclear Cardiology Protocols Cross-Sectional Study (INCAPS). Eur Heart J 2015; 36: 1689–96.

6. Suzuki Y, Matsumoto N, Makita A, Ashida T, Kuronuma K, Amano Y, et al. Hybrid assessment of myocardial ischemia using stress-only nuclear myocardial perfusion imaging and rest computed tomography perfusion imaging. Circ J 2020; 84: 1818–25.

7. Ronan G, Wolk MJ, Bailey SR, Doherty JU, Douglas PS, Hendel RC, et al. ACCF/AHA/ASE/ASNC/HFSA/HRS/SCAI/SCCT/SCMR/STS 2013 multimodality appropriate use criteria for the detection and risk assessment of stable ischemic heart disease: a report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, American Heart Association, American Society of Echocardiography, American Society of Nuclear Cardiology, Heart Failure Society of America, Heart Rhythm Society, Society for Cardiovascular Angiography and Interventions, Society of Cardiovascular Computed Tomography, Society for Cardiovascular Magnetic Resonance, and Society of Thoracic Surgeons. J Nucl Cardiol 2014; 21: 192–220.

8. Hesse B, Tägil K, Cuocolo A, Anagnostopoulos C, Bardiés M, Bax J, et al. EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology. Eur J Nucl Med Mol Imaging 2005; 32: 855–97.

9. Henzlova MJ, Cerqueira MD, Mahmarian JJ, Yao SS, Quality Assurance Committee of the American Society of Nuclear Cardiology. Stress protocols and tracers. J Nucl Cardiol 2006; 13: e80–90.

10. Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ. Radiation dose to patients from cardiac diagnostic imaging. Circulation 2007; 116: 1290–305.

11. Ueyama T, Takehana K, Maeba H, Iwasaka T. Prognostic value of normal stress-only technetium-99m myocardial perfusion imaging protocol. Comparison with standard stress-rest protocol. Circ J 2012; 76: 2386–91.

12. Gerber BL, Belge B, Legros GJ, Lim P, Poncelet A, Pasquet A, et al. Characterization of acute and chronic myocardial infarcts by multidetector computed tomography: comparison with contrast-enhanced magnetic resonance. Circulation 2006; 113: 823–33.

13. Cousins C, Miller DL, Bernardi G, Rehani MM, Schofield P, Vañó E, et al. ICRP PUBLICATION 120: Radiological protection in cardiology. Ann ICRP 2013; 42: 1–125.

14. Henzlova MJ, Duvall WL, Einstein AJ, Travin MI, Verberne HJ. ASNC imaging guidelines for SPECT nuclear cardiology procedures: Stress, protocols, and tracers. J Nucl Cardiol 2016; 23: 606–39.

15. Zafrir N, Bental T, Solodky A, Ben-Shlomo A, Mats I, Hassid Y, et al. Feasibility of myocardial perfusion imaging with half the radiation dose in obese patients using ordered-subset expectation maximization with resolution recovery software. J Nucl Cardiol 2013; 20: 111–9.

16. Nakazato R, Berman DS, Hayes SW, Fish M, Padgett R, Xu Y, et al. Myocardial perfusion imaging with a solid-state camera: simulation of a very low dose imaging protocol. J Nucl Med 2013; 54: 373–9.

17. Perrin M, Djaballah W, Moulin F, Claudin M, Veran N, Imbert L, et al. Stress-first protocol for myocardial perfusion SPECT imaging with semiconductor cameras: high diagnostic performances with significant reduction in patient radiation doses. Eur J Nucl Med Mol Imaging 2015; 42: 1004–11.

 
© The Japanese Society of Nuclear Cardiology 2022
feedback
Top