板状誘電体 を用いたジャイロトロン RF出力強度分布の定量的測定
真鍋 亜佑斗田畑 邦佳髙瀬 芳貴森 映樹小紫 公也関根 北斗辻 政裕小田 靖久
著者情報
ジャーナル フリー

2022 年 30 巻 2 号 p. 51-58

詳細
抄録

A new method using a thick dielectric plate is discussed to measure the intensity distribution of a gyrotron's RF output in one step, unlike the conventional method in two steps. 1D simulation of the thermal diffusion in the plate removes the effect of convection cooling from the measured temperature distribution on the backside of the plate. As a result, the output radiation of UT-94 gyrotron is successfully measured. The peak intensity and the integrated power are 0.048 GW/m2 and 64.5 kW , respectively, within the uncertainty of 2.6%. The error of the measurement is expected to be less than 8.0 %.