Quantitative measurement of high-intensity focused ultrasound fields at various optical propagation distances and driving voltages using background-subtracted shadowgraph method
Yuma Ono, Shohei Mori, Shin Yoshizawa
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: e26.114

Details
Abstract

Conventional hydrophones for high-intensity focused ultrasound (HIFU) measurement suffer from invasiveness and damage risks. To address these challenges, in this study, a background-subtracted shadowgraph method enabled quantitative, non-contact visualization of HIFU fields. The accuracy was verified using a fiber-optic hydrophone, demonstrating a relative error within 10% up to 10.1 MPa. Additionally, varying the optical propagation distance revealed a trade-off between the measurable pressure range and the signal-to-noise ratio. Furthermore, stable measurements were achieved in the low-pressure range (1 MPa) with only four averaged frames. These results firmly establish the quantitative reliability of this non-contact optical method.