Article ID: e26.114
Conventional hydrophones for high-intensity focused ultrasound (HIFU) measurement suffer from invasiveness and damage risks. To address these challenges, in this study, a background-subtracted shadowgraph method enabled quantitative, non-contact visualization of HIFU fields. The accuracy was verified using a fiber-optic hydrophone, demonstrating a relative error within 10% up to 10.1 MPa. Additionally, varying the optical propagation distance revealed a trade-off between the measurable pressure range and the signal-to-noise ratio. Furthermore, stable measurements were achieved in the low-pressure range (1 MPa) with only four averaged frames. These results firmly establish the quantitative reliability of this non-contact optical method.