Annals of Vascular Diseases
Online ISSN : 1881-6428
Print ISSN : 1881-641X
ISSN-L : 1881-641X
ORIGINAL ARTICLES
Jotec E-Ventus BX Stent Graft Deployment in the FEVAR and Iliac Branch Device: Single Centre Experience
Tamer Sayed Islam AhmedAlexander RodwayKarim El SakkaSyed Waquar Yusuf
著者情報
ジャーナル オープンアクセス

2019 年 12 巻 2 号 p. 171-175

詳細
抄録

Objectives: To evaluate the outcomes of the E-ventus BX balloon-expandable stent graft system (Jotec, Hechingen, Germany) implanted as bridging stent grafts during fenestrated endovascular aortic repair (FEVAR) and the iliac branch device (IBD) of complex aneurysms.

Methods: This was a single centre retrospective analysis prospective study including all consecutive patients treated by FEVAR and the IBD performed with E-ventus BX stent grafts as bridging stents. Demographics of patients, the diameter and length of the bridging stent grafts, technical success, reinterventions, occlusions, post-operative events, and imaging (computed tomography [CT] scan and ultrasound) were prospectively collected in an electronic database. Follow-ups were performed with clinical assessment and a CT angiogram scan at four weeks after discharge followed by a duplex ultrasound every six months for two years and then a yearly duplex scan afterwards.

Results: Between June 2015 and October 2017, 40 consecutive patients (three females) were treated with custom made fenestrated endografts and the iliac branch device for complex aneurysms, using the E-Ventus BX stent graft. All 82 E-Ventus BX stent grafts were successfully delivered and deployed. There was no in-hospital mortality. The early bridging stents patency rate was 97.6% (80 out of 82). The two-target vessel post-operative occlusion was secondary to kink of the renal stents and failure for re-lining of the renal artery. Of the two patients, only one needed permanent dialysis. On the late follow-up (after 30 days), two other patients demonstrated a renal stent occlusion, with one treated successfully with re-lining of the stent and the other patient treated conservatively. Neither of them needed permanent dialysis. A follow-up was maintained for 36 patients until April 2018 with a median follow-up of 18 months. All bridging stents E-Ventus BX stent grafts remained patent (78 out of 82, 95.1%).

Conclusion: E-Ventus BX stent grafts used as bridging stents during FEVAR and the IBD are associated with favourable outcomes at the mid-term follow-up. Long-term follow-up is required to confirm these promising results.

著者関連情報
© 2019 The Editorial Committee of Annals of Vascular Diseases. This article is distributed under the terms of the Creative Commons Attribution License, which permits use, distribution, and reproduction in any medium, provided the credit of the original work, a link to the license, and indication of any change are properly given, and the original work is not used for commercial purposes. Remixed or transformed contributions must be distributed under the same license as the original.

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 継承 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-sa/4.0/deed.ja