One Technique to Modulate a Device Implantation Path in a Short Treatment Length Using the Gore IBE Device
Hiroshi Mitsuoka Yuta MiyanoYasuhiko TeraiShinnosuke GotoShinji KawaguchiMasanao NakaiFumio Yamazaki
Author information
JOURNAL OPEN ACCESS
Supplementary material

2019 Volume 12 Issue 3 Pages 408-411

Details
Abstract

The GORE EXCLUDER Iliac Branch Endoprosthesis (IBE; W. L. Gore and Associates, Flagstaff, AZ, USA) applicability is limited by the aorto-iliac length (AOL). The shortage may be a major exclusion criterion. An 85-year-old male presented with an abdominal aortic and left common iliac arterial aneurysm. The left-side AOL was 146-mm, which was deemed 19-mm too short for IBE usage. To increase implantation length, the contra-lateral connection stent graft was deployed along the implantation line, wound half-circumferentially around the ipsilateral limb. Any form of endoleak, limb occlusion, and device migration has not been observed for twelve months.

Content from these authors
© 2019 The Editorial Committee of Annals of Vascular Diseases. This article is distributed under the terms of the Creative Commons Attribution License, which permits use, distribution, and reproduction in any medium, provided the credit of the original work, a link to the license, and indication of any change are properly given, and the original work is not used for commercial purposes. Remixed or transformed contributions must be distributed under the same license as the original.

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 継承 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-sa/4.0/deed.ja
Previous article Next article
feedback
Top