Annals of Vascular Diseases
Online ISSN : 1881-6428
Print ISSN : 1881-641X
ISSN-L : 1881-641X
Case Report
Successful Treatment of Type 3b Endoleak after AFX Using TREO
Tobuhiro NitaHironori Baba Yuji HironakaGen ShinoharaYoshie OchiaiShigehiko Tokunaga
Author information
JOURNAL OPEN ACCESS

2025 Volume 18 Issue 1 Article ID: cr.25-00016

Details
Abstract

Late postoperative type 3 endoleaks associated with the AFX (Endologix, Irvine, CA, USA) stent-graft are commonly reported. The AFX’s endoskeletal design raises concerns about wire entrapment between its frame and fabric, as well as a risk of type 1a endoleak. An 84-year-old man with prior EVAR using the AFX presented with a type 3b endoleak 4 years postoperatively. He underwent relining with the TREO stent-graft (Terumo Aortic, Sunrise, FL, USA), which enabled easy contralateral gate cannulation and secured a long proximal landing zone. The TREO appears to be a viable option for relining the AFX in type 3b endoleak cases.

Content from these authors
© 2025 The Editorial Committee of Annals of Vascular Diseases

This article is licensed under a Creative Commons [Attribution 4.0 International] license.
https://creativecommons.org/licenses/by/4.0/
Previous article Next article
feedback