2025 Volume 18 Issue 1 Article ID: nmt.25-00069
During thoracic endovascular aneurysm repair for aortic arch aneurysms, deployment of the stent graft parallel to the aortic neck is crucial to preventing a type Ia endoleak from the proximal end. We report the early tip capture release and push-up technique that comprises early release of the proximal bare stent, which is typically deployed last during stent graft deployment, followed by a push-up maneuver after landing the proximal edge, thus allowing conformation to the aortic morphology. This technique is effective even for complex aortic arch anatomy.