Annals of Vascular Diseases
Online ISSN : 1881-6428
Print ISSN : 1881-641X
ISSN-L : 1881-641X
Case Reports
Migration of the Zenith Flex Device during Endovascular Aortic Repair of an Infrarenal Aortic Aneurysm with a Severely Angulated Neck
Yukihisa OgawaHiroshi NishimakiKiyoshi ChibaKenji MurakamiYuka SakuraiKeishi FujiwaraTakeshi MiyairiYasuo Nakajima
Author information
JOURNAL FREE ACCESS

2016 Volume 9 Issue 3 Pages 232-234

Details
Abstract

A woman in her 80s with an infrarenal aortic aneurysm was scheduled for endovascular aortic repair (EVAR). The aneurysm had a severely angulated neck (SAN), and the Zenith Flex device was selected. Completion angiography showed migration of the main body resulting in right renal artery stenosis. A Palmaz genesis was placed across the renal orifice. The patient had no renal dysfunction and was discharged 7 days after EVAR. If Zenith Flex devices are used for a SAN, it is necessary to consider not only the position of the renal artery but also the appropriate position of the stent-graft.

Content from these authors
© 2016 Annals of Vascular Diseases

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 継承 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-sa/4.0/deed.ja
Previous article Next article
feedback
Top