Balloon-trapped Technique for Successful OptEase Filter Retrieval
Katsuhiro KondoHirotoshi WatanabeMasashi IwabuchiMasakiyo Nobuyoshi
著者情報
ジャーナル フリー 早期公開

論文ID: cr.10.01041

この記事には本公開記事があります。
詳細
抄録
Purpose: To describe a case of successful retrieval of OptEase filter, using the balloon-trapped technique, after failure of its retrieval by the standard techniques.
Case report: An 82-year-old man had an inferior vena cava filter placed for deep venous thrombosis prophylactically. Seven days after successful catheter-directed thrombolysis, the filter was retrieved after changing the position to caudal, using a standard angioplasty balloon. The patient was discharged on warfarin anticoagulant.
Conclusion: Balloon-trapped removal of the OptEase filter that failed standard retrieval attempt is an effective and safe technique that can be performed using commonly available tools and familiar to most interventionalists.
著者関連情報
© 2011 Japanese College of Angiology, Japanese Society for Vascular Surgery, Japanese Society of Phlebology