Annals of Vascular Diseases
Online ISSN : 1881-6428
Print ISSN : 1881-641X
ISSN-L : 1881-641X

This article has now been updated. Please use the final version.

Collapse and Occlusion of Balloon-Expandable Covered Stent in the Right Common Iliac Arteries Due to External Stress
Seimei Go Taira KobayashiTaiichi TakasakiShinya Takahashi
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: cr.24-00038

Details
Abstract

The patient previously underwent surgery for cervical cancer. She reported intermittent claudication, and computed tomography (CT) revealed total occlusion of the left iliac artery. We conducted endovascular treatment (EVT) using balloon-expandable covered stents (BECS). The patient underwent colostomy closure after EVT. The patient reported intermittent claudication beginning the day after the procedure. The CT showed a collapse of the distal side of the stent implanted in the right common iliac artery, accompanied by localized thrombo-occlusion. During open surgery, metal hooks typically affect the iliac artery; force transmission through the tissue may indirectly compress the iliac artery, leading to BECS failure.

Content from these authors
© 2024 The Editorial Committee of Annals of Vascular Diseases

This article is licensed under a Creative Commons [Attribution-NonCommercial-ShareAlike 4.0 International] license.
https://creativecommons.org/licenses/by-nc-sa/4.0/