Noninvasive Technique for Measuring Central Venous and Arterial Pressure Using Controlled Compression Sonography
Hiroshi Tomoeda Kentaro SawadaShingo Chihara
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: oa.20-00058

Details
Abstract

Objective: Devices that can noninvasively measure central and peripheral venous pressures with relative ease and in a short time were developed, but the resolution of the data that can be recorded with these devices is limited to 50 mmHg.

Materials and Methods: We aimed to develop a system that could overcome this limitation. We used an innovative noninvasive controlled compression sonography device that could theoretically measure pressures higher than 200 mmHg. First, to validate the accuracy of our device, an in vitro study was conducted. Then, the values measured by our system were compared to conventionally obtained measurements of central venous, peripheral venous, and brachial artery pressures. Finally, regression analyses were used to determine the correlations between measurements obtained from different devices.

Results: With our device, the measurement of venous and arterial pressures required only 3 to 15 sec. All regression analyses revealed a significant statistical correlation between measurements, although the correlation coefficient was relatively low for arterial pressure.

Conclusion: For venous pressure, our system can provide measurements that could not be measured noninvasively with conventional methods. Regarding arterial pressure, although our system could measure systolic pressure, further studies are needed to confirm the clinical efficacy of our device.

Content from these authors
© 2020 The Editorial Committee of Annals of Vascular Diseases. This article is distributed under the terms of the Creative Commons Attribution License, which permits use, distribution, and reproduction in any medium, provided the credit of the original work, a link to the license, and indication of any change are properly given, and the original work is not used for commercial purposes. Remixed or transformed contributions must be distributed under the same license as the original.