抄録
The accuracy of analytical value of C in steel by QUANTOVAC analysis depends on the purity of argon gas.After some experiments, the cause of unsteadiness of C value in the analysis was found to be hydrocarbon in argon gas.Use of a special gas bomb provided with a molecular sieve lter led to satisfactory results.