分光研究
Online ISSN : 1884-6785
Print ISSN : 0038-7002
ISSN-L : 0038-7002
Y字形反応管を用いた水銀の原子吸光分析
白浜 和夫渡辺 誠治真木 武
著者情報
ジャーナル フリー

1972 年 21 巻 5 号 p. 316-321

詳細
抄録

In the determination of mercury by atomic absorption Spectrometory, the high sensitivity is obtained by the addition of reducing agents.But the decrease of absorbance is observed on standing.To eliminate this standing effects, Y-type reaction tube was used.The process is that mercury solution and reducing agent solution are aspirated separately and the two solutions are mixed in a Y-type reaction tube.Furthermore, the effect of reducing agents, acids, pH and diverse ions were investigated.Among reducing agents, tin (II) chloride was found to be the best agent and absorbance could be raised to about fifteen-folds.Detection limit of this method was 0.02ppm of mercury and the coefficient of variation was 1.1% at 10ppm Hg level. When the ammonium nitrate was contained, the absorbance of mercury was constant at a pH of 1-10.
One ppm of I- and 200ppm of SCN- affected, but interference of other ions is negligible. The analytical results of mercury in numerical indicator tube by this method was satisfactorily in agreement with dithizone method.

著者関連情報
© 社団法人 日本分光学会
前の記事 次の記事
feedback
Top