2015 年 49 巻 4 号 p. 239-246
Inductively coupled plasma (ICP) is characterized by high ionization efficiency for almost all elements. Recently, a triple quadrupole ICP-MS (ICP-MS/MS) has been applied to the measurements of isotope ratios for ultra-low level radioisotopes. The ICP-MS/MS features an additional quadrupole mass filter situated in front of the octopole reaction cell and quadrupole mass filter, which allows only the analyte ion to enter the cell by rejecting all the other ions. In this manuscript, we review recent studies on environmental radioactivity in Fukushima achieved by ICP-MS/MS techniques.