Does Size Matter? Cryoballoon-Based Pulmonary Vein Isolation Using a Novel 25-mm Circular Mapping Catheter
Bruno ReissmannMichael SchlüterFrancesco SantoroTilman MaurerChristian-Hendrik HeegerChristine LemesThomas FinkJohannes RiedlAndreas RilligShibu MathewFeifan OuyangKarl-Heinz KuckAndreas Metzner
Author information
JOURNAL FREE ACCESS FULL-TEXT HTML Advance online publication

Article ID: CJ-17-0956

Details
Abstract

Background: Real-time recording of pulmonary vein isolation (PVI) using a circular mapping catheter has become a key aspect of cryoballoon (CB) ablation. The aim of this study was to investigate the procedural safety, efficacy and rate of real-time pulmonary vein (PV) recording using a novel circular mapping catheter with a 25-mm loop size for CB-based PVI.

Methods and Results: A total of 40 patients with symptomatic atrial fibrillation (AF) underwent PVI using a second-generation CB and a novel 25-mm circular mapping catheter. A total of 159 PV were identified and successfully isolated. Real-time PV recording was achieved in 80% of the PV. In 3 right inferior PV the circular mapping catheter had to be exchanged for a stiff guidewire due to insufficient mechanical support. Therefore, acute PVI using exclusively the circular mapping catheter was achieved in 156/159 PV (98%). Mean procedure and fluoroscopy times were 66±21 min and 15±6 min, respectively. Transient phrenic nerve palsy occurred in 1 patient as the only procedural complication.

Conclusions: The exclusive use of a novel 25-mm circular mapping catheter for CB ablation of AF results in a real-time PV recording rate of 80% and isolation of 98% of targeted PV.

Cryoballoon (CB) ablation is an effective treatment option for paroxysmal and persistent atrial fibrillation (AF).17 The procedural endpoint is the electrical isolation of the pulmonary veins (PV),8 assessed using circular mapping catheters with loop diameters of 15 or 20 mm incorporating 8 electrodes (Achieve, Arctic Front Advance, Medtronic, Minneapolis, MN, USA). Initial ablation strategies were based on fixed freeze cycle durations of 180–240 s, frequently also including an added bonus freeze application of the same duration.1,4,9,10 Real-time PV recordings gained further importance since a correlation of the time to isolation (TTI) of the PV with arrhythmia-free survival was demonstrated.11,12 Therefore, novel ablation strategies take the individual TTI into account when setting the total application time, aiming at shorter freeze cycles and thus at reduction of potential complications such as phrenic nerve (PN) palsy and esophageal thermal injury.1214 The consistent application of these TTI-guided ablation protocols, however, is hampered by a limited real-time verification of PV isolation (PVI) when using currently available 15- or 20-mm diameter circular mapping catheters with 8 electrodes.1114

Recently, a newly designed circular mapping catheter with loop diameters of 15, 20 and 25 mm and a modified solid-core shaft was introduced (Figure 1). The latter is equipped with 10 evenly spaced electrodes (Achieve AdvancedTM, Medtronic). The aim of this study was therefore to assess the rate of real-time recordings from inside the PV, acute efficacy, and procedural safety during the routine use of the 25-mm mapping catheter.

Figure 1.

(A) First-generation 20-mm circular mapping catheter with its 8 electrodes and (B) the novel 25-mm circular mapping catheter incorporating 10 electrodes.

Methods

Inclusion and Exclusion Criteria

Patients with symptomatic paroxysmal or persistent AF were included in the current study. Exclusion criteria were prior left atrial (LA) ablation, LA diameter >60 mm, severe valvular heart disease or contraindications to post-interventional oral anticoagulation.

Periprocedural Management

Transesophageal echocardiography was performed prior to PVI in all patients to rule out intracardiac thrombi and to assess the LA diameter. No further pre-procedural imaging was performed. Ablation was performed under a therapeutic international normalized ratio (INR) of 2–3. Novel oral anticoagulants were stopped the day before the procedure and continued 6 h after ablation. All patients provided written informed consent. The study is based on a retrospective analysis and was approved by the local ethics board and was performed in accordance with the Declaration of Helsinki.

Ablation

Patients were placed under deep sedation using i.v. sufentanil, midazolam and propofol. Two 8-F sheaths (St. Jude Medical, Minneapolis, MN, USA) were placed in the right femoral vein. A 6-F steerable decapolar catheter (InquiryTM, St. Jude Medical) was positioned in the coronary sinus. A single trans-septal puncture was performed via the right femoral vein under fluoroscopy using a modified Brockenbrough technique and an 8.5-F trans-septal sheath (SL1, St. Jude Medical). Heparin was given to maintain an activated clotting time ≥300s. Selective angiographic visualization by injection of contrast medium was performed to identify the individual PV ostia. The trans-septal sheath was then exchanged over a guidewire for a 12-F steerable sheath (FlexCath AdvanceTM, Medtronic) through which the 28-mm CB was advanced into the LA. Guiding of the CB to the target PV was performed over a 25-mm inner lumen circular mapping catheter (Achieve AdvancedTM, Medtronic), and complete occlusion of the PV ostium was verified by contrast injection through the central lumen of the inflated CB. The circular mapping catheter facilitated real-time verification of PVI by recordings from the PV ostium, allowing for an ablation strategy taking the individual TTI into account.13 When PVI was verified, freezing was continued for an additional fixed period of 120 s.13 If the TTI could not be recorded, the freeze cycle duration was set to 180 s. No additional bonus freeze cycle was applied after successful PVI.

An esophageal temperature probe (Sensitherm, St. Jude Medical; or Circa, Circa Scientific) was inserted and positioned according to the individual CB position to provide esophageal temperature monitoring during energy delivery. The intraluminal esophageal temperature cut-off was set at 15℃.15

During delivery of cryoenergy along the septal PV, continuous PN pacing was performed using a diagnostic catheter positioned in the superior vena cava (6F; InquiryTM, St. Jude Medical). PN capture was monitored by tactile feedback of diaphragmatic contraction when the operator’s hand was placed on the patient’s abdomen. In addition, the continuous motor action potential (CMAP) was monitored and refrigerant delivery was stopped immediately if weakening or loss of diaphragmatic movement was noted or the amplitude of the CMAP decreased by 30%.16,17

Post-Procedure Care

Transthoracic echocardiography was performed in all patients to rule out pericardial effusion. All patients were treated with proton-pump inhibitors for 6 weeks. Low-molecular-weight heparin was administered to patients on vitamin K antagonists with INR <2.0 until therapeutic INR 2–3 was achieved. Novel oral anticoagulants were reinitiated 6 h after ablation. Anticoagulation was continued for ≥3 months, and thereafter based on the individual CHA2DS2-VASc score. Previously ineffective anti-arrhythmic drugs were continued for 3 months.

Endpoints

The primary endpoint was the rate of real-time recordings of CB-based PVI documented by the 25-mm circular mapping catheter. Secondary endpoints included (1) electrical isolation of all PV confirmed on circular-mapping catheter recordings from the PV; (2) procedural parameters (e.g., procedure duration, TTI); and (3) major complications (transient ischemic attack, stroke, pericardial tamponade, PN palsy, and severe bleeding requiring blood transfusion).

Statistical Analysis

All data were evaluated retrospectively. Continuous data are described as mean±SD if normally distributed; otherwise as median (IQR, first quartile; third quartile). Categorical data are described with absolute and relative frequencies. All calculations were performed using SAS (version 9.3, SAS Institute, Cary, NC, USA).

Results

Patient Characteristics

A total of 40 consecutive patients with symptomatic paroxysmal or persistent AF underwent 28-mm second-generation CB-based PVI at Asklepios Klinik St. Georg, Hamburg. In all procedures, the 25-mm circular mapping catheter was used for guidance of the CB and real-time verification of PVI. Patient baseline characteristics are listed in Table 1.

Table 1. Patient Characteristics
No. patients 40
Age (years) 60±13
Female 13 (33)
BMI (kg/m2) 28±5
LA diameter (mm) 44±5
LVEF (%) 55±4
CHA2DS2-VASc score 2 (1–2)

Data given as mean±SD, median (IQR) or n (%). BMI, body mass index; LA, left atrium; LVEF, left ventricular ejection fraction.

Procedural Parameters

In the 40 patients, a total of 159 PV were identified (40 right superior PV [RSPV], 40 right inferior PV [RIPV], 39 left superior PV [LSPV], 39 left inferior PV [LIPV] and 1 left common PV [LCPV]; Table 2). All PV were eventually successfully isolated. With the first CB application, electrical PVI was achieved in 38 RSPV (95%), 30 RIPV (75%), 28 LSPV (72%), 32 LIPV (82%), and 1 LCPV (100%). The mean number of freeze cycles resulting in PVI was 1.1±0.3, 1.5±1, 1.3±0.6, 1.2±0.5 and 1 for the RSPV, RIPV, LSPV, LIPV, and LCPV, respectively. Real-time recording of PVI was achieved in 35 RSPV (88%), 30 RIPV (75%), 28 LSPV (72%), 33 LIPV (85%), and in 1 LCPV (100%). The overall rate of real-time PV recordings was 80% (95% CI: 73–86%).

Table 2. Acute Ablation Results
  RSPV RIPV LSPV LIPV LCPV
No. PV 40 40 39 39 1
Isolated PV 40 (100) 40 (100) 39 (100) 39 (100) 1 (100)
No. applications until PVI 1.1±0.4 1.5±1.0 1.3±0.6 1.2±0.5 1
Total freeze cycle duration (s) 149±27 174±41 172±26 159±25 165
Freeze-cycle duration until confirmed PVI (s) 25 (20–30) 41 (29–52) 44 (32–50) 29 (23–41) 45
No. PVI confirmed in real time 35 (88) 30 (75) 28 (72) 33 (85) 1 (100)
Min. balloon temperature (℃) −52±5 −50±5 −48±5 −45±5 −45
Min. endoluminal esophagus temperature (℃) 35±1 32±6 35±1 34±2 21

Data given as mean±SD, median (IQR) or n (%). Three RIPV could not be isolated using the 25-mm circular mapping catheter; in each case, exchange of the mapping catheter for a stiff guidewire resulted in successful cryoballoon-based PVI thereafter. LCPV, left common PV; LIPV, left inferior PV; LSPV, left superior PV; PV, pulmonary vein; PVI, PV isolation; RIPV, right inferior PV; RSPV, right superior PV.

Three RIPV, each characterized by a low insertion level into the LA, could not be ablated and isolated using the 25-mm mapping catheter due to incomplete PV occlusion despite multiple balloon and mapping catheter maneuvers and despite different sheath angulations. Even when aiming at maximum mechanical support by forgoing real-time electrical recordings but positioning the mapping catheter at distal levels of the target PV, occlusion was never perfect. The mapping catheter was then replaced for a stiff guidewire (Amplatz Super StiffTM Guidewire, Boston Scientific, Plymouth, MN, USA) aiming at more mechanical support. In each of these cases, successful CB-based PVI was then achieved. Thus, the rate of PVI achieved using exclusively the 25-mm mapping catheter was 156/159 (98%; 95% CI: 95–100%).

Median freeze cycle duration until real-time verification of PVI was 25 s (IQR, 20–30 s) for RSPV, 41 s (IQR, 29–52 s) for RIPV, 44 s (IQR, 32–50 s) for LSPV, 29 s (IQR, 23–41 s) for LIPV, and 45 s for the LCPV (Table 2). Mean total freeze cycle duration including PV without real-time signal recordings and fixed freeze cycle durations of 180 s was 149±27 s, 174±41 s, 172±26 s, 159±25 s, and 165 s for the RSPV, RIPV, LSPV, LIPV and LCPV, respectively. Mean minimum balloon temperature and mean minimum endoluminal esophageal temperature for the respective PV are also given in Table 2. Mean procedure and fluoroscopy times were 66±21 min and 15±6 min, respectively. The median dose area product was 637 cGycm2 (IQR, 436–929 cGycm2).

25-mm Circular Mapping Catheter: Special Aspects

In most PV the 25-mm circular mapping catheter can be easily advanced into the target PV and positioned in close proximity to the balloon surface in its normal circular shape, thereby providing signals from all 10 electrodes. In contrast, when the 25-mm mapping catheter is used in small PV or in PV with multiple proximal branches, insertion of the catheter into one of these branches often resulted in a characteristic reversed circular mapping catheter position (Figure 2). In this position, the proximal electrodes were deep inside the PV and the distal electrodes close to the PV orifice, allowing for efficient mechanical support and resulting in PV signal recordings only from the distal electrodes. In some of these cases, however, the distal electrodes of the mapping catheter were trapped between the balloon surface and the antral PV tissue, making the real-time differentiation between far-field atrial and PV potentials more challenging (Figures 2,3A). Additionally, the close proximity of these electrodes to the CB surface resulted in freezing artifacts in an early phase of the respective freeze cycle, thereby preventing a clear real-time assessment of PVI (Figure 3B).

Figure 2.

Complete ostial occlusion of the right inferior pulmonary vein (RIPV) is verified by injection of contrast medium (CM) through the central lumen of the inflated cryoballoon (CB). Note the reversed circular mapping catheter (SMC) position. The proximal electrodes are deep inside an inferior branch of the RIPV while the distal electrodes are in an ostial position and in this example trapped in between the balloon surface and the antral pulmonary vein tissue. DC, diagnostic catheter; ETP, esophageal temperature probe; TS, trans-septal sheath. Fluoroscopic right anterior oblique 30° view.

Figure 3.

(A) Tracing of surface electrocardiogram (ECG) leads I, II, and V1 and intracardiac electrograms recorded from the 10 electrodes of the 25-mm circular mapping catheter (PV inf 1–10) within the right inferior pulmonary vein (RIPV; cf. Figure 2). A diagnostic catheter is positioned in the superior vena cava to perform continuous phrenic nerve pacing (CS 1–2). Real-time signals from the circular mapping catheter are recorded only by distal electrodes (PV inf 1–2 and PV inf 3–4; arrows), which are in an antral position, whereas the proximal electrodes are deep inside the RIPV. The electrograms from the proximal electrodes reflect fusion of far-field atrial and near-field PV potentials. (B) Surface ECG leads I, II, and V1 and intracardiac electrograms recorded from the 10 electrodes of the 25-mm circular mapping catheter (PV inf 1–10) positioned within the right inferior pulmonary vein (cf. Figure 1). Continuous phrenic nerve pacing is performed by a diagnostic catheter positioned in the superior vena cava (CS 1–2). The close proximity of the distal circular mapping catheter electrodes (PV inf 1–2 and PV inf 3–4) to the surface of the cryoballoon results in freezing artifacts starting at 24 s of the freeze cycle, thus preventing a clear assessment of real-time recording of PV isolation (arrows).

Complications

Transient PN palsy was observed in 1 patient during CB ablation along the RSPV; at that time, the RSPV had already been successfully isolated. Freezing was immediately stopped and the PN recovered within 30 min. No further complications occurred.

Discussion

The current study is the first to report procedural characteristics using a novel 25-mm circular mapping catheter routinely for CB ablation of AF. Major findings are (1) rate of real-time PV recording of 80%; (2) high acute success rates; and (3) high procedural safety.

Real-Time PV Recording

The first-generation inner lumen circular mapping catheter allows for over-the-wire support of the balloon, real-time PV recordings, and assessment of PVI. Currently the CB in combination with a circular mapping catheter is the only balloon-based ablation system that allows for routine and systematic real-time assessment of PV signals. First studies evaluating mapping technologies in conjunction with the hot-balloon, however, are currently under investigation. The TTI independently predicts recovered PV-LA conduction11 and thereby directly affects arrhythmia-free survival.12,18 The direct feedback on freeze cycle efficacy enables the operator to anticipate ineffective freeze cycles, which should result in premature discontinuation of the respective application. In contrast, challenging variants of PV or LA anatomies might prevent PVI at an early stage of the freeze cycle; in this setting, freeze cycle durations of 180 s or 240 s might be too short to ensure durable PVI, and individualized freeze cycle durations >180 s or 240 s might be necessary. Therefore, an increased incidence of real-time PV recordings will facilitate individualized ablation strategies and might further enhance the efficacy and safety of CB-based PVI.12,13

The rate of real-time recordings using the first-generation inner lumen circular mapping catheter was in the range of 40–76%.13,19 In the current study, the use of the novel second-generation 25-mm circular mapping catheter resulted in a rate of real-time PV recordings of 80%. In 3 of the present patients, however, the circular mapping catheter had to be exchanged for a stiff guidewire after failed attempts at adequate PV occlusion due to insufficient mechanical support. The use of a stiff guidewire then resulted in complete PV occlusion and successful PVI in all 3 patients. This is in contrast to previous publications describing successful isolation of all targeted PV using exclusively the first-generation 20-mm circular mapping catheter.13,14,19 The requirement to replace the 25-mm mapping catheter in these patients might be attributed to the new solid-core shaft design of the catheter, which appears to be less stiff than the previous first-generation circular mapping catheter; it thereby potentially offers less mechanical support in certain anatomical settings, for example, a low insertion level of the RIPV into the LA and early branching in close proximity to the LA-PV junction. In this particular setting, we think that the stiffer shaft of the first-generation Achieve catheter provides better mechanical support and stability than the new design shaft of the second-generation Achieve catheter. All other PV were successfully isolated using standard CB maneuvers.

Learning Curve

The fluoroscopy times in the current study are longer compared with a previous publication on CB ablation from our group.14 The performance, however, of the new circular mapping catheter with its new shaft design is different to the previous catheter generation, and requires a certain learning curve even when procedures are performed by experienced operators.

Study Limitations

The current study is an observational, non-randomized single-center analysis with a limited number of patients, and more experience is required before final conclusions on the new circular mapping catheter generation can be drawn.

Conclusions

The exclusive use in a routine clinical setting of a novel 25-mm circular mapping catheter for CB ablation of AF resulted in a real-time PV recording rate of 80% and isolation of 98% of targeted PV. TTI-guided ablation protocols might be effectively supported.

Acknowledgments

None.