COMPARISON OF THE CARDIAC OUTPUT MEASUREMENTS WITH ELECTROMAGNETIC FLOWMETER AND DYE-DILUTION TECHNIQUE IN DOGS, WITH A NOTE ON THE COMPARISON BETWEEN RADIOCARDIOGRAPHY AND EAR-DYE METHOD IN MAN
SENRI HIRAKAWAHIROYASU ITOYASUSHI KONDOSHIGEKI KANYAMAIKUO WATANABENORIYASU MORISHOJI HAYASE
著者情報
ジャーナル フリー

1975 年 39 巻 4 号 p. 393-402

詳細
抄録
There are relatively few studies and opinions are still divided. as to the agreement between the electromagnetic flowmeter and dye-dilution technique in measurements of cardiac output in animals. Of 22 pairs of cardiac output measurements performed with an electromagnetic flow probe applied to the root of aorta in open-chest, anesthetized dogs and with peripheral venous injection of indocyanine green dye and an arterial cuvette, the measured points fell within the range of ± 20% on both sides of the line of identity in 19 instances. Studied on 8 human subjects with radiocardiography and ear-dye method performed simultaneously with peripheral venous injection of a mixture of R131 ISA and indocyanine green, cardiac output given by ear-dye method was higher than that given by radiocardiography by an average of 9 ± 11 (SD)%. Agreement and disagreement in cardiac output measurements given by electromagnetic flow probe, dye-dilution method with arterial cuvette, dye-dilution method with ear-piece densitometer and radiocardiography were discussed.