JAPANESE CIRCULATION JOURNAL
Online ISSN : 1347-4839
Print ISSN : 0047-1828
ISSN-L : 0047-1828
Evaluation of Cardiac Motion and Function by Cine Magnetic Resonance Imaging : ADVANCES IN IMAGING TECHNIQUES FOR DIAGNOSIS OF CARDIOVASCULAR DISEASES
TAKESHI KONDOHIROSHI KUROKAWAHIROFUMI ANNOMASAHIRO OKAMURANAOKO ANNOKAZUYOSHI SAKAKURAMAMORU TOKUDAHITOSHI HISHIDAYOSHIHIKO WATANABEYASUSHI MIZUNOMUNETAKA SUGIISHI
著者情報
ジャーナル フリー

1992 年 56 巻 6 号 p. 632-638

詳細
抄録

Cardiac cine magnetic resonance imaging (MRI) was studied to evaluated the cardiac motion and function, and a water-stream phantom study was performed to clarify whether it was possible to quantitatively assess the valvular regurgitation flow by the size of the flow void. In normal subjects, the left ventricular (LV) epicardial apex swung up to the base only a few millimeters, and the mitral annulus ring moved about 14 mm as mean value toward the apex during systole. Those motions of mitral annulus ring may contribute to the left atrial filling. The LV longitudinal shortening and torsions were shown by the tagging method. This tagging method was the best method for estimating cardiac motions. Cardiac cine MRI using software including a modified Simpson's method program and a wall motion analysis program was useful for routine LV volumetry and wall motion analysis because it was a simple and reliable method. Our water-stream phantom studies demonstrated that it might be difficult to perform quantitative evaluation of valvular regurgitation flow by using only the size of the flow void without acquiring in-formation relating to the orifice area.

著者関連情報
© Japanese Circulation Society