A case of amyotrophic lateral sclerosis managed by tracheostomy and invasive ventilation in which air leaks occurred at the cuff
Nobuhiko ShibasakiKaoru KonishiYutaka NishiyamaTetsuo MiyagawaTakaya Numayama
Author information
JOURNAL OPEN ACCESS Advance online publication
Supplementary material

Article ID: cn-001990

Details
Abstract

The patient was a 64-year-old woman who had been diagnosed with amyotrophic lateral sclerosis 8 years ago, and had been under artificial ventilation with tracheotomy for 6 years. Computed tomography indicated a dilated tracheal diameter of 29.6 ‍mm at the cuff, and a high cuff pressure of 80 ‍cmH2O. An adjustable flange tracheostomy tube with an optional length setting was used to extend the effective length by 28 ‍mm. A previously evident air leak disappeared with the change in cuff level, and cuff pressure decreased to 25 ‍cmH2O. X-ray images indicated a reduction in the size of the previous cuff area. Tracheal dilatation due to improper management of cuff pressure is a contributing factor to air leakage at the cuff area, and using an adjustable flange tracheostomy tube in an effort to resolve such air leaks is a valid option.