ウエアラブル化学センサと応用に関する研究
三林 浩二, 若林 慶彦, 井口 滋仁, 岡本 敏明
著者情報
ジャーナル 認証あり

2002 年 4 巻 5 号 p. 5_1-5_6

詳細
抄録
Wearable oxygen and conductimetric sensors were constructed by microfabrication techniques for monitoring biochemical substances. The wearable oxygen sensor (thickness 84 μm) was calibrated against dissolved oxygen from 0 to 8.0 mg/l, and was successfully used to monitor the transcutaneous oxygen pressure by applying the device onto the forearm skin surface. Tear fluid conductivity was possible to be measured by applying the wearable sensor to the human subject’s temporal lower cul-de-sac, thus evaluating electrolyte concentration and turnover rate in tear for normal, healthy volunteers and dry-eye patients.