電気学会論文誌D(産業応用部門誌)
Online ISSN : 1348-8163
Print ISSN : 0913-6339
ISSN-L : 0913-6339
心臓ペースメーカ充電のための経皮的エネルギー伝送システム
西村 敏博江口 哲治井上 猛斎藤 正男前島 靖真嶋 政文平地 克也
著者情報
ジャーナル フリー

1997 年 117 巻 9 号 p. 1085-1091

詳細
抄録
Cardiac pacemaker has to be changed in three to five years for the battery life. A patients must be performed an operation whenever a patient is changed cardiac pacemaker. The method employing secondary rechargeable battery is researched to extend the battery life. The secondary rechargeable battery is transcutaneously charged from outside the body. This paper presents a transcutaneous energy transmission system for a rechargeable cardiac pacemaker battery. The transcutaneous energy transmission system is constructed with a high frequency resonant inverter, a transcutaneous transformer and a rectifier circuit. The secondary transformer is piled and the coupling coefficient and the conversion efficiency are measured with an experimental set up. The results show as follows. The best diameter of the primary center core is 15[mm]. The best output voltage frequency is 10[kHz].
著者関連情報
© 電気学会
前の記事 次の記事
feedback
Top