Case Reports
When a Burr Can Not Penetrate the Calcified Lesion, Increasing Burr Size as Well as Decreasing Burr Size Can Be a Solution in Rotational Atherectomy
Kenichi SakakuraYousuke TaniguchiKei YamamotoHiroshi WadaShin-ichi MomomuraHideo Fujita
Author information
JOURNAL FREE ACCESS

2017 Volume 58 Issue 2 Pages 279-282

Details
Abstract

In rotational atherectomy (RA), several burr sizes are available, such as 1.25 mm, 1.5 mm, 1.75 mm, or ≥ 2.0 mm. It is important to select an appropriate burr size for each lesion because rotational atherectomy has several unique complications regarding burrs such as entrapment or perforation. When a burr cannot penetrate the lesion, downsizing of the burr is generally recommended. Also, if the smallest burr (1.25 mm) cannot penetrate the lesion, a change to a more supportive or larger French guiding catheter has been recommended. We describe the case of a 68 year-old female who was referred to our department for percutaneous coronary intervention to the calcified stenosis in the middle of the left anterior descending coronary artery. We used the smallest burr (1.25 mm) and a supportive 7 Fr guiding catheter to penetrate the lesion. However, the smallest burr could not pass the lesion even after 14 sessions (total ablation time: 339 seconds). We intentionally increased the burr size from 1.25 mm to 1.5 mm. The 1.5 mm burr successfully passed the lesion without any perforation or burr entrapment. In this manuscript, we discuss why increasing the burr size was successful for this severely calcified lesion that was not penetrated by the smallest burr.

Content from these authors
© 2017 by the International Heart Journal Association
Previous article Next article
feedback
Top