Clinical Influence and Predictors of Pacing-Induced Mechanical Asynchrony in Patients with Normal Cardiac Function with Ventricular Lead Placed in Non-Apical Position
Analysis Based on Intrinsic Conduction Disturbance
Masato OkadaKazunori KashiwaseAkio HirataYasuharu TakedaRyohei AmiyaYasunori UedaYoshiharu HiguchiYoshio Yasumura
著者情報
ジャーナル フリー
電子付録

2018 年 59 巻 6 号 p. 1275-1287

詳細
抄録

Right ventricular apical (RVA) pacing often causes left ventricular (LV) mechanical asynchrony, which is enhanced by impaired cardiac contraction and intrinsic conduction abnormality. However, data on patients with normal cardiac function and under RV non-apical (non-RVA) pacing are limited.

We retrospectively investigated 97 consecutive patients with normal ejection fraction who received pacemaker implantation for atrioventricular block with the ventricular lead placed in a non-RVA position. We defined mechanical asynchrony as discoordinate contraction between opposing regions of the LV wall evaluated by echocardiography. Asynchrony was detected in 9 (9%) patients at baseline and in 38 (39%) under non-RVA pacing (P < 0.001). Asynchrony at baseline was significantly associated with complete left bundle branch block (CLBBB) [odds ratio (OR) = 20.8, P < 0.001]. Asynchrony under non-RVA pacing was significantly associated with left anterior fascicular block (LAFB) (OR = 7.14, P < 0.001) and CLBBB (OR = 13.3, P = 0.002) at baseline. New occurrence of asynchrony was significantly associated with LAFB at baseline (OR = 5.88, P = 0.001). During a median follow-up period of 4.8 years, the incidence of device-detected atrial fibrillation (AF) was more frequent in patients who developed asynchrony than in those who did not (53.3% versus 27.5%, hazard ratio = 2.17, 95% confidence interval = 1.02-4.61, P = 0.03).

In patients with normal cardiac function, LAFB at baseline was significantly associated with new occurrence of mechanical asynchrony under non-RVA pacing. Abnormal contraction had a significant influence on the incidence of device-detected AF.