International Heart Journal
Online ISSN : 1349-3299
Print ISSN : 1349-2365
ISSN-L : 1349-2365
Case Reports
A Steam Pop Detected by Intracardiac Echocardiography During Catheter Ablation of the Left Ventricular Papillary Muscle
Akinori SatoNobue YagiharaYuki HasegawaSou OtukiKenichi IijimaDaisuke IzumiTohru Minamino
Author information
JOURNAL FREE ACCESS

2019 Volume 60 Issue 1 Pages 199-203

Details
Abstract

A 60-year-old female with premature ventricular contractions (PVCs) originating from the bottom of the posteromedial papillary muscle of the left ventricle underwent radiofrequency catheter ablation (RFCA) using an irrigated-tip catheter. During ablation of the PVCs, a loud steam pop was observed. Intracardiac echocardiography (ICE) revealed a growing, hyperechogenic intramyocardial microbubble formation around the catheter tip. The formation disappeared slowly and completely, leaving an endocardial laceration without pericardial effusion. ICE imaging is valuable during a difficult RFCA procedure, because ICE reveals the exact anatomical position of the catheter and thus allows rapid evaluation of the occurrence of steam popping and any possible subsequent complication.

Content from these authors
© 2019 by the International Heart Journal Association
Previous article Next article
feedback
Top