2026 Volume 67 Issue 6 Pages 536-543
Very high-power short-duration (VHPSD) ablation has been proposed to produce shallower lesions due to a predominance of resistive over conductive heating, potentially reducing collateral tissue injury. However, in vivo data on luminal esophageal temperature (LET) during VHPSD pulmonary vein isolation (PVI) remain limited.
A total of 107 consecutive patients (62 ± 10 years; 75 male; 88 with paroxysmal atrial fibrillation) underwent PVI using VHPSD (90 W/4 second) at the posterior wall and high-power short-duration (HPSD; 50 W, ablation index 450) at the anterior wall. LET was continuously monitored during the procedure, and all sites with LET elevation ≥ 39°C were analyzed.
LET ≥ 39°C occurred in 41 patients (38.3%), with a mean peak LET of 40.05 ± 0.99°C. The mean time to peak LET was 13.9 seconds, while temperature returned to within 1°C of baseline after 65 seconds following cessation of radiofrequency (RF) delivery. LET elevations ≥ 39°C occurred at distances of 3.76 ± 1.73 mm in the anteroposterior (AP) view and 2.83 ± 1.10 mm in the left anterior oblique (LAO) view from the temperature sensors. Notably, 89% of sites with LET ≥ 39°C were associated with the left inferior pulmonary vein (LIPV).
VHPSD ablation is associated with LET elevations ≥ 39°C, particularly during ablation near the inferior pulmonary veins. Peak temperature occurs after completion of RF delivery, limiting real-time modulation of application duration. Given the prolonged temperature decay, rapid adjacent RF applications should be avoided. LET monitoring should be considered during VHPSD ablation.