Dosimetry of Occupational Eye Lens Dose Using a Novel Direct Eye Dosimeter, DOSIRIS, during Interventional Radiology Procedures
Masakazu HirakawaHiroshi NakatakeSatoru TsurutaShuji MatsuuraYuushi MotomuraYoshiki HirakiKoshi MimoriKousei Ishigami
Author information
JOURNAL OPEN ACCESS

2022 Volume 7 Issue 2 Pages 40-43

Details
Abstract

In response to the recommendation by the International Commission on Radiological Protection to lower the equivalent eye dose limit, the Japanese Government in April 2021 lowered the equivalent dose limit for the eye lens for occupational exposure. A considerable number of interventional radiology operators are exposed to levels above the new limit. For this reason, a need exists to more accurately evaluate eye lens dose in interventional radiology operators by using a novel direct eye dosimeter, the DOSIRIS™ (IRSN, France), which is capable of measuring a 3-mm dose equivalent under protective glasses. The DOSIRIS is a thermoluminescent dosimeter that exhibits good energy dependence and better directional properties than other dosimeters. Dosimetry using DOSIRIS might be accurate and compatible with the latest regulations.

Content from these authors
© 2022 Japanese Society of Interventional Radiology