Interventional Radiology
Online ISSN : 2432-0935

This article has now been updated. Please use the final version.

Percutaneous Retrieval of a Misinserted Pigtail Catheter Straightener during Infra-Renal Abdominal Aortic Stenting: A Case Report
Yuichiro IzumiShuji IkedaAkira KitagawaYusuke ImaedaTetsuya YamadaMakiyo HagiharaToyohiro OtaTsuneo IshiguchiKojiro Suzuki
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: 2020-0018

Details
Abstract

Percutaneous retrieval of an intravascular foreign body is a minimally invasive technique. Using cone-beam computed tomography and the lateral grasp technique, we successfully retrieved a pigtail catheter straightener that had been misinserted into the right common iliac artery. Some examples of catheter straightener retrieval have been reported; however, it is important to take care not to accidentally insert a catheter straightener into a vessel via an angiographic sheath.