Untangling a Knotted Angiographic Catheter Using a Balloon Catheter: A Case Report
Motonori TakahagiHaruyuki TakakiHiroyuki YokoyamaJunichi TaniguchiHisashi KomotoAtsushi OgasawaraHiroshi KodamaYasukazu KakoKaoru KobayashiKoichiro Yamakado
著者情報
ジャーナル オープンアクセス 早期公開

論文ID: 2022-0036

この記事には本公開記事があります。
詳細
抄録

When a 66-year-old man with hepatocellular carcinoma underwent an angiographic examination, a 4-Fr catheter was inserted from the right femoral artery. It became tightly knotted in the descending aorta. To untangle the knotted catheter, a noncompliant balloon catheter was delivered into the knotted loop from the contralateral femoral artery. After the balloon catheter was inflated from the inside of the knotted loop, the knot became loose. Finally, the knotted catheter was untangled. Subsequently, the remainder of the examination was performed as initially planned.