Development of High-resolution TV Camera for Real Time Digital Radiography System
Yoichi OnoderaHisatake YokouchiMasayuki TsuneokaFumitaka Takahashi
Author information
JOURNAL FREE ACCESS

1993 Volume 47 Issue 5 Pages 732-738

Details
Abstract
We have developed a 2000-line TV camera which use a 1-inch SATICON as a real-time digital radiography (DR) system
We record a 60% amplitude response when 1000 TV lines are at the center of the image, and a 46 dB signal-to-noise ratio using the 2100 line scanning mode at a signal current of 550 nA