Abstract
A novel sampling strategy suitable for compressed sensing MRI (CS-MRI) is proposed. Spatial resolution of
reconstructed images is sometimes reduced due to variable density random undersampling. To improve spatial resolution in CSMRI,
oblique scanning in k-space for the read-out direction is adopted to acquire a higher spatial frequency for the phaseencoding
direction. Reconstruction experiments showed that spatial resolution can be improved without lengthening scan time.