88. 減速材付き線量計による中性子線量当量測定の標準化とその移行法の検討
小佐古 敏荘中村 尚司道川 太一加藤 和明中沢 正治辻本 忠木村 逸郎関口 晃
著者情報
ジャーナル フリー

1982 年 24 巻 2 号 p. 139-149

詳細
抄録

Using the primary neutron standard in the Electrotechnical Laboratory, the moderating-type neutron Rem-Counter with internal absorber (Studsvik 2202D type) in the Institute for Nuclear Study of the University of Tokyo was evaluated as a secondary standard detector by measuring its energy response function. By combining this energy response function with that obtained at PhysikalischTechnische Bundesanstalt in West Germany, the value of neutron dose equivalent was evaluated based on this secondary standard detector. Furthemore, the same type detectors of the six labolatories in Japan were intercompared in the neutron fields formed by 252Cf and 241Am-Be neutron sources, and the standardized value of neutron dose equivalent was transferred to these detectors. Some discussions have been carried out about problems in transferring and using this value of neutron dose equivalent in order to establish the traceability of neutron standard.