Measurement of Radon-222 Concentration in Environment Sampled within Short Time Using Charcoal Detector
Tadashi YAMASAKIShigenobu SEKIYAMAMina TOKINYumiko NAKAYASUTamaki WATANABE
Author information
JOURNAL FREE ACCESS

1994 Volume 36 Issue 8 Pages 752-758

Details
Abstract
The concentration of 222Rn in air sampled within a very short period of time was measured using activated charcoal as the adsorber. The detector is the plastic canister containing mixture of the activated charcoal and the silica gel. The radon gas was adsorbed in the charcoal in the radon chamber at the temperature of 25°C. A little amount of liquid scintillation cocktail was added into the vial of liquid scintillation counter with the canister. The radon in the charcoal was extracted in the liquid scintillation cocktail. Alpha particles emitted from radon and its daughter nuclei in the cocktail were detected using the liquid scintillation counter. Present method has advantages of not only short sampling time of air but also adsorption of radon in charcoal under a constant temperature. The concentration of radon in air down to 2Bq/m3 could be detected. A kinetic model for adsorption of radon in the charcoal is also presented. The ratio of radon concentration in the charcoal to that in air under the equilibrium state of adsorption was estimated to be from 6.1 to 6.8m3/kg at the temperature of 25°C.
Content from these authors