日本原子力学会誌
Online ISSN : 2186-5256
Print ISSN : 0004-7120
ISSN-L : 0004-7120
遠隔ピペッター
江村 悟中原 嘉則大貫 守河野 信昭
著者情報
ジャーナル フリー

1967 年 9 巻 12 号 p. 705-709

詳細
抄録

Two positive delivery pipetters (1 ml and 5ml in capacity) have been devised for pipetting aliquots of highly radioactive liquid samples intended, various analytical deteminations in the reprocessing analytical facility at the Japan Atomic Energy Research Institute.
The pipetter assembly consists of pipet, electrical elevator, control unit, three valves and trap, as well as nitrogen gas and vacuum lines.
The structural material of the parts of the pipet in contact with the sample are either stainless steel or teflon.
A precision of 0.2% standard deviation of successive 100μl deliveries of distilled water and concentrated nitric acid was obtained on the 1 ml pipetter.
Design criteria for the pipetter, its principle of operation and method of decontamination are described.

著者関連情報
© by the Atomic Energy Society of Japan
前の記事 次の記事
feedback
Top